+86 19963093538
Current location:HOME>PRODUCT>Micro Spectrophotometer>Nucleic Acid and Protein Concentration Analyzer
Nucleic Acid and Protein Concentration Analyzer

Nucleic Acid and Protein Concentration AnalyzerYP-CWF2


【Introduction】:Youyunpu Optoelectronic Tech specializes in R&D & production of Nucleic Acid and Protein Concentration Analyzer, with custom services. Complete Nucleic Acid and Protein Concentration Analyzer models, high detection accuracy, trusted by customers.

Product Details

Product Introduction

The ultra-micro spectrophotometer is a high-precision biological detection instrument based on ultraviolet-visible spectrophotometry (UV-Vis). It is specially designed for the rapid analysis of concentration and purity of biological samples such as nucleic acids (DNA/RNA), proteins and cell lysates with extremely low sample volumes (0.5–2 μL).

Detection Principle

Based on the Beer-Lambert Law (A=εcl) — the absorbance (A) of a solution is proportional to the extinction coefficient (ε), path length (c) and concentration (l).

Application Scope

The ultra-micro spectrophotometer is widely used in life science research, biopharmaceuticals, medical diagnosis, food safety and other fields.

Key Advantages

  1. ★ Android 10 operating system, supporting data upload via Wi-Fi and Ethernet for fast batch data uploading.

  2. 7-inch high-definition Android touch screen with 1024×600 pixels and clear, efficient UI interface, enhancing operation and visual experience.

  3. ★ Equipped with a supervision platform; test results can be directly uploaded to the platform for remote data monitoring.

  4. ★ Built-in 7000mAh high-capacity battery, providing up to 8 hours of continuous operation on a full charge, ensuring stable operation anytime and anywhere. Lightweight and portable design enables easier mobile operation.

  5. Adopts liquid column surface tension technology to realize rapid detection of ultra-micro samples of 0.5–2 μL, without cuvettes or capillaries, significantly reducing sample consumption.

  6. No sample dilution required; the measurable concentration range is more than 150 times that of conventional UV-Vis spectrophotometers.

  7. Adopts unique motor control technology and three-path detection mode, with better stability, repeatability, linearity and wider detection range.

  8. Consumable-free design allows samples to be wiped off directly or recovered by pipette after measurement, with simple and efficient workflow, especially suitable for precious, scarce or volume-limited biological samples.

  9. Single detection takes only a few seconds from sample application to result output, suitable for high-throughput rapid screening and routine quality control.

  10. Compared with traditional detection methods, the light source flashing algorithm extends light source life, reduces light intensity stimulation, enables faster detection and less degradation of samples.

  11. Equipped with OD600 optical detection system and cuvette detection module, which can directly measure the concentration of bacterial culture, expanding applications in microbiology research and cell culture quality control.

  12. Fluorescence detection module combined with fluorescent quantitative kits, through specific binding of fluorescent dyes to target substances, reduces the DNA/RNA detection limit to 0.5 pg/μL (dsDNA), realizing accurate quantification of trace samples.

  13. Light source fluctuation pp value (peak intensity minus minimum intensity) < 0.5%.

  14. Built-in high-performance fiber optic spectrometer.

  15. High-stability pulsed xenon light source.

  16. User-defined model function.

  17. Built-in new-generation high-speed thermal printer.

  18. Data stored in image and table formats.

Technical Specifications

  1. Wavelength range: 230 nm, 260 nm, 280 nm (three wavelengths)

  2. Sample volume: 0.5–2 μL

  3. Path length: 0.2 mm, 0.04 mm (high-concentration measurement), 1.0 mm (normal-concentration measurement)

  4. Light source: xenon flash lamp

  5. Detector: UV-enhanced CMOS linear array sensor

  6. Nucleic acid detection range: 25–11000 ng/μL (dsDNA)

  7. Absorbance precision: 0.003 Abs (0.2 mm path length)

  8. Absorbance accuracy: ±1% (7.332 Abs @ 260 nm)

  9. Absorbance range: 0.5–220 A

  10. Detection time: < 6 s

  11. OD600 absorbance range: 0–6 Abs

  12. OD600 absorbance stability: (0, 3) ≤ 0.5%; (3, 4) ≤ 2%

  13. OD600 absorbance repeatability: (0, 3) ≤ 0.5%; (3, 4) ≤ 2%

  14. Fluorescence detection linearity: R² > 0.995

  15. Fluorescence detection repeatability: < 1.5%

  16. Fluorescence detection stability: < 1.5%

  17. Internal data storage: 10,000–100,000 records, queryable and exportable via USB

  18. Power adapter: 45 W fast charging

  19. Battery capacity: 21.6 V 7000 mAh

  20. Power consumption: 25 W (standby: 10 W)

  21. Dimensions: 325 × 227 × 211 mm

  22. Weight: 5.15 kg

Related Products