+86 19963093538
Current location:HOME>PRODUCT>Specialized Instruments>GMO Detector
GMO Detector

GMO DetectorYP-P16


【Introduction】:Youyunpu Optoelectronic Tech specializes in R&D & production of GMO Detector, with custom services. Complete GMO Detector models, high detection accuracy, trusted by customers.

Product Details

Instrument Overview

The Agricultural Product GMO Detector is a device used to identify the presence of genetically modified components in agricultural products. Genetically modified agricultural products are crops into which foreign genes have been introduced via genetic engineering technology. This detector enables agricultural producers, food processing factories, regulatory authorities and other stakeholders to perform rapid detection and identification of genetically modified ingredients.

Detection Procedure

The GMO Detector operates based on molecular biological techniques, with the workflow consisting of the following steps:
  1. Sample Extraction
    Extract DNA or RNA samples from agricultural products, which contain the genetic information of the tested material.
  2. PCR Amplification
    Amplify specific gene fragments associated with transgenic events using polymerase chain reaction (PCR) technology.
  3. Detection Method
    Various techniques can be used to analyze the amplified products, including gel electrophoresis, DNA microarray analysis, real‑time fluorescent PCR, etc., to confirm the presence of transgenic events.
  4. Data Analysis
    Analyze and interpret the detection results to determine whether the sample contains genetically modified components.
The use of the GMO Detector helps producers and relevant authorities ensure product quality and safety, and maintain compliance with applicable regulations and standards. It also supports informed consumer choices and improves food information transparency.

Device Specifications

ItemSpecification
Mechanical
Dimensions (W×D×H)235 mm × 385 mm × 175 mm
Weight5.6 kg
Electrical
Power~220 V / 50 Hz, 255 W
Data Interface2 × USB 2.0 (right side)
Environmental
Operating Conditions10–30 ℃, 20%–80% RH
Storage & Transport−20–55 ℃, 20%–80% RH
Altitude< 2500 m
NoiseA‑weighted, < 60 dB
Sample
Sample Capacity16 × 0.2 mL
Tube TypeSingle tubes, 8‑tube strips, 16‑well blocks
Reaction Volume15–100 μL
Thermal Performance
Heating/CoolingSemiconductor heating / cooling
Temperature Range4 ℃ – 99 ℃
Max Heating Rate≥ 3.5 ℃/s
Avg Heating Rate≥ 2.5 ℃/s
Max Cooling Rate≥ 3.5 ℃/s
Avg Cooling Rate≥ 2.5 ℃/s
Temperature Precision≤ ±0.01 ℃
Temperature Accuracy≤ ±0.1 ℃
Temperature Uniformity≤ ±0.3 ℃
Optical Performance
Fluorescence Channels2
Light SourceHigh‑brightness LED
DetectorHigh‑sensitivity, high‑SNR photodiode
Supported Probes/DyesChannel 1: 470/520 – FAM, SYBR Green
Channel 2: 530/570 – HEX, JOE, VIC
Detection Sensitivity1 copy
Linear Range10⁰ – 10¹⁰ copies
Correlation Coefficient≥ 0.999
Channel CrosstalkNone
Repeatability≤ 1.0%